The Solution of Water-free handpiece with Multi-point Water Spray

Release time:

2023-12-28 09:46

The multi-point water spray handpiece  suddenly has less water output or no water output, which is generally caused by the following reasons:
1. The handpiece  uses tap water, and the external water pressure is small;
2. The waterway of handpiece  is blocked;
3. The atomizing gas of the handpiece   is too large, and the difference between the atomizing air pressure and the water pressure is too large. The air forces the water back and the water cannot come out. Or although the atomizing air switch has not been moved, the water pressure has decreased, and the atomizing air has forced the water to stop and the water cannot come out.
We can through the following steps to troubleshoot and solve:
1. First try to adjust the water quantity switch of the handpiece  and the atomizing air switch of the handpiece  to see if the water can be discharged normally. Generally on the right side of the tool tray, there are handpiece  water switch andhandpiece   atomization air switch. handpiece  water switch some brand dental chair is 3, some is 1. handpiece  atomization air switch, usually 1.
(1) Turn off the handpiece  atomization air switch, that is, the atomization air is adjusted to the minimum.
(2) Adjust the water outlet switch of the handpiece  , transfer the water out of the handpiece  , and adjust the water output to the appropriate size.
(3) Re-open the handpiece  atomization air switch, and gradually increase the atomization air to obtain the ideal atomization effect.
(4) If the problem can be solved through the above 3 steps, the handpiece  does not need to be repaired. The next time you encounter a similar problem, you can use this method to solve it.
(5) if there is no water after repeated adjustment, it is necessary to check whether there is water at the joint of the handpiece .
2. Take down the handpiece  , step on the foot switch, and check whether there is water at the joint of the handpiece  pipe.
① If there is water out of the handpiece  connector, check the water storage bottle, water storage bottle switch, water source changeover switch and handpiece  water volume switch in turn. Make sure that there is water in the water storage bottle, the water storage bottle switch is turned on, the water source switch is in the pure water position, and the handpiece  water volume switch is turned on.
② After the operation above ①, the handpiece  connector still has no water, which can be determined as the fault of the dental chair. Please contact the after-sales service department of the dental chair.
③ After the operation above ①, if there is water at the joint of the handpiece  , it can be determined that the waterway of the handpiece  is blocked and returned to the factory for repair.











 