Maintenance and Maintenance of Dental Handpiece
Release time:
2023-12-28 09:57
Handpiece cleaning and disinfection have the following taboos:
1. Can not use ultrasonic cleaning machine to wash handpiece . The cleaning vibration frequency of the ultrasonic cleaning machine 30000 times per minute will destroy the retainer of the handpiece bearing, causing the handpiece bearing to fall apart (as shown in the right figure), and the whole handpiece will be scrapped.
2. It is strictly prohibited to clean handpiece and surgical instruments together. Impurities, fibers and other substances attached to surgical instruments can cause damage to the handpiece when they enter the handpiece during cleaning.
3.It is strictly forbidden to use cleaning enzyme and disinfection enzyme to clean the handpiece . If the enzyme substance is not fully cleaned, there will be residue in the handpiece bearing, and the crystallized crystal will stick the steel ball or ceramic ball of the bearing and the retainer together. The bearing will not turn. The handpiece no longer works.
4. It is strictly prohibited to soak the burs with aldehydes or enzymes. The handpiece burs should also be sterilized with a three-time vacuum sterilizer. If it is soaked with aldehydes or phenols, it will be picked up from the soaking liquid when it is used. The residual soaking liquid will follow the burs into the handpiece head, and the bearing will be locked after crystallization, so the handpiece cannot be rotated.

The cleaning and maintenance of handpiece, according to actual situation, there are many methods, in general, have to go through the following processes:
Cleaning-drainage of cleaning residue-oiling-drainage of residual oil -bagging -sanitizing-storage
Now introduce three common methods:
1.Use a professional medical cleaning machine (cleaning and maintenance all-in-one machine) for cleaning, oiling, maintenance and drying. After the completion of the maintenance and then sealing, sterilization, preservation.

2.Use a professional medical cleaning machine for cleaning; idle rotation in the turbine with compressor motor or dental chair to discharge the cleaning residual liquid in the handpiece ; automatic tanker machine with oil ; Bag sealing; Sterilization; Preservation.

3.Use cleaning lubricant to clean handpiece users. Turn the handpiece in the clear water for 5 - 10 seconds to remove the stains in the handpiece . Cover the air return pipe of the handpiece and spray the handpiece cleaning lubricant into the air inlet pipe for 3 - 5 times. Idling the turbine and draining excess cleaning lubricant. Bagging, sterilization, preservation.

Contact Information
Address: Room 502, Block 9, Liandong Youzhi Park, No. 10 Senshuji Road, Nansha Community, Danzao Town, Nanhai District, Foshan City
Contact Phone:0757-85411586/81263523
Mailbox:tenry@fsqiyang.com
Leave Us A Message.
Contact Us
Contact Phone:0757-29269893
Official website:www.shieldtex.cn
Contact Email:qiuwq@shieldtex.cn
Add: No.2, Shizhou Section, Baichen Road, Gangbei Industrial Zone, Shizhou Village, Chencun Town, Shunde District, Foshan City
OUR PRODUCTS
Focus on us
Copyright©2023 Guangdong Shishi Medical Technology Co., Ltd.